Understanding The ISO/TS
16949:2002 Standard
7.6 Control of Monitoring and
Measuring Devices (MMD)
Determine what
monitoring and measurement is required and what MMD devices are
needed to provide evidence of product conformity (see
7.2.1)
Establish
processes to ensure that monitoring and measurement can
be carried out and is in fact being carried out in a manner
that is consistent with requirements.
When necessary to
ensure valid results:
a)
Calibrate or verify MMD at specified intervals, or prior to
use, against measurement standards that are traceable to
national or international standards. Record the basis for
calibration if no such standard exists.
b)
Adjust or re-adjust MMD as necessary
c)
Identify MMD so that its calibration status can be
determined
d)
Safeguard MMD from improper adjustments that would invalidate
the measurement result
e)
Protect MMD from damage and deterioration during handling,
maintenance and storage
Assess and record the
validity of previous measuring results if when the MMD does not
meet requirements
Take appropriate action
on such MMD and any product affected
Keep records of the
calibration and verification of MMD’s
Confirm the ability of
computer software to satisfy the intended application, when it
is used to measure and monitor to specified
requirements.
Confirm this ability prior to initial use and reconfirm as
necessary.
Note:
See ISO 10012-1 and 10012-2 for
guidance
Note:
A number or other
identifier traceable to the device calibration
record meets the
intent of 7.6c.
7.6.1 Measurement System
Analysis
Conduct statistical
studies to analyze variation present in the results of each
type of MMD that is referenced in the Control
Plan.
Use analytical
methods & acceptance criteria
that:
Conform to methods
and criteria in customer reference (MSA) manuals Or use other
methods, if approved by the
customer
7.6.2
Calibration/verification
Records
Keep records of the
calibration or verification of company & employee owned
MMD’s to show evidence of product conformity. These records
shall include:
Þ
MMD identification, including the measurement standard used to
calibrate the MMD
Þ
Revisions following engineering
changes
Þ
Any out-of-specification readings as received for calibration
or verification
Þ
An evaluation of the impact of the out-of-specification
condition
Þ
Statement of conformity to the specification after calibration
or verification
Þ
Notification to the customer if suspect product or material
been shipped
7.6.3 Laboratory
Requirements
7.6.3.1 Internal
Laboratory
In your QMS
documentation, define the scope for your laboratory that
includes:
Þ
It’s capability to perform required inspection, test &
calibration services
Your laboratory
shall specify & implement technical requirements
that, as a minimum, include:
Þ
Adequacy of laboratory procedures
Þ
Competency of laboratory personnel
Þ
Product testing
Þ
Capability to perform these services correctly and be traceable
to relevant process standards (such as ASTM, EN,
etc,)
Þ
Review of related quality records
Note
– your internal laboratory may be accredited to ISO/IEC 17025,
but it is not mandatory.
7.6.3.2 External/commercial/independent Labs
Where you use an
external, commercial or independent laboratory for inspection,
test or calibration services, it shall have a defined
laboratory scope that includes its capability to perform the
inspection, test or calibration services you
require.
You shall obtain
evidence that the external laboratory
is:
Þ
Accredited
to ISO/IEC 17025 or national
equivalent
Þ
OR
- that it is acceptable to the customer
Note 1:
Evidence of customer acceptance may include an audit by the
customer or an audit by a customer-approved 2nd
party that the laboratory meets the intent of ISO/IEC
17025
Note 2:
The equipment manufacturer of a specific MMD, can perform its
calibration, if a qualified laboratory is not available. In
such cases, you shall ensure that clause 7.6.3.1 requirements
are met.
Key Explanation Points and
Tips:
Þ
Requirements for what needs to be measured (see clause 4.1c
& e; 7.1c) and the acceptance criteria (see clause 7.2.1;
7.1c; 7.3.3c) may come from the customer, regulatory, industry
and your own organization. Product realization planning (see
clause 7.1) must determine the following – what specific
product and process characteristics needs to be monitored and
measured; the criteria for product acceptance; the type of
monitoring and measurement device needed; frequency - at what
stages of realization to do it; sample size; etc.
Þ
You must then determine what MMD is appropriate for each
measuring or monitoring requirement. Consideration must be
given to the measurement capability (precision) of the MMD
which may have to be several times greater than the tolerance
criteria for product measurement. This would depend on the
industry you are in and the criticality of end use for the
product (e.g. the precision requirements for an engine block or
for ball bearings may be much greater than say for cutting
leather to cover a car seat).
Þ
Personnel using MMD’s must have competence and training in
the use of MMD’s in terms of their function, range and
precision of measurement, reliability, use and maintenance.
Þ
MMD’s may include measurement and testing tools; equipment;
hardware and software. They may be owned by your organization;
your employees or the customer. MMD’s may be used to verify
product as well as to measure process conformity (e.g. a
temperature controller on an oven). Besides MMD’s used for
product conformity, you may need to calibrate and control
certain MMD’s used in related and peripheral processes such as
production equipment; tooling; maintenance; etc.
Þ
To ensure valid measurement and monitoring results, MMD’s must
be controlled. A process is required, to control the
selection; purchase; identification; status; calibration; use;
verification or adjustment; use; handling; maintenance and
storage; training; nonconforming MMD’s; calibration records;
etc. Appropriate records need to be kept of the use of these
controls.
Þ
All MMD’s used for product verification must be capable of
being calibrated, verified or both. Calibration is setting or
correcting an MMD, usually by adjusting it to match or conform
to a dependably known and traceable standard (e.g. adjusting a
micrometer or caliper to conform to master blocks traceable to
national standards).
Þ
Verification is confirming that the MMD is meeting or
performing to acceptable national measurement standards and
does not involve any correction or adjustment (e.g. verifying a
ruler or tape measure against a calibrated ruler that has been
calibrated to a national standard). A ruler or tape measure is
generally not capable of being calibrated and when it gets out
of calibration its use must be discontinued.
Þ
There are MMD’s that are capable of being both calibrated and
verified (e.g. a CMM- coordinate measuring machine) and may
require both to be done in specific situations based on
frequency of use and criticality of measurement. This
requirement also applies to the use of computer software whose
calibration status must be established prior to initial use and
reconfirmed (verified) at defined intervals.
Þ
You must define the frequency and method of calibration for
each type and level (shop floor; laboratory or standard) of
MMD. Your calibration records must identify what standard you
used for calibration and show traceability of the standards you
use at your facility to national or international standards.
Þ
In rare circumstances, national or international standards may
not exist for calibrating a specific MMD. In such situations
consider using industry, manufacturer or even your own
organizational standard to validate the accuracy and
reliability of your MMD. Consult with your customer if the
contractual circumstances require it.
Þ
Your control plan must define the measurement and monitoring
required and the type of MMD needed for it, including the
frequency of measurement and acceptance criteria. Use customer
reference manuals, such as the Measurement Systems Analysis
(MSA) manual, to conduct statistical studies on MMD’s
referenced in your control plans. Ensure that personnel
performing such statistical studies are trained and competent
to do so.
Þ
A multitude of software tools are available to manage and
control MMD’s including all the record keeping details required
by clause 7.6.2. There are many acceptable methods to identify
MMD’s and their calibration status. The methods you select must
consider the manufacturers recommendations; frequency of use;
environment the MMD is used in; etc.
Þ
Where an MMD is found to be out of calibration, you must take
appropriate correction action to contain and re-verify the
product affected, to the extent practical. This is in addition
to containing, repair and recalibration of the defective MMD.
Þ
Customer or internal engineering changes may result in a change
in product measurement, requirements and/or the MMD to be
used. These
changes would normally be reflected in your control plan.
Ensure that your calibration process shows clear linkage to
your process for change control (clause 7.1.4) and control plan
(clause 7.5.1.1).
Þ
The quality of measurement data is subject to
variability related to the measuring device, the measuring
process, the operator using the measuring device, the product
being measured, the environment the measurements are made in,
etc. The study and control of the statistical
characteristics (bias, repeatability, reproducibility,
stability and linearity) that measure these variables is called
Measurement System Analysis
(MSA).
Þ
The TS 16949 standard requires that the MMD or category
(verniers, calipers, etc.) of MMD referenced in product Control
Plans be subject to statistical analysis. The analysis methods
and acceptance criteria for the statistical characteristics
referred to above must conform to automotive OEM - MSA
reference manuals. Other methods and acceptance criteria may be
used if approved by the customer. Also review specific MSA
requirements at OEM customer or IATF
websites.
Þ
In many organizations, the internal laboratory may conduct more
technical and comprehensive inspection, testing and calibration
using more complex and sensitive equipment, methods and
standards. You must have document the internal laboratory
scope; You must also specify technical requirements for –
adequacy of procedures; personnel training and competency;
testing methods; traceability to relevant process standards;
control of test specimens; records needed, etc.
Þ
Your internal laboratory scope must specify the tests,
evaluations and calibrations it is qualified to perform;
provide a list of the equipment used to perform these
activities; and a list of the methods, standards, etc.,
used.
Þ
The procedures used in the laboratory could be established
practices; MMD manufacturer’s reference or user manuals;
industry standards, methods and practices; customer specified
methods; and regulatory methods and practices. These procedures
typically address testing methods and standards; identification
and traceability; etc. The need to have a documented laboratory
procedure or manual would depend on the scope and complexity of
product testing and inspection.
Þ
OEM customers may have specific competency and training
requirements for laboratory personnel. Refer to OEM websites;
IAOB reference material or contact your OEM customer for the
latest requirements.
Þ
If you use an external laboratory, you must have evidence that
it is ISO/IEC 17025 (or national equivalent) accredited or
acceptable to the customer. Ensure that the external
laboratory’s ISO/IEC 17025 accreditation is not out of date and
its scope includes the activities you have contracted it to
perform. Also review specific external laboratory requirements
at OEM customer or IATF websites.
Þ
Clause 4.1f requires you to continually improve your processes.
Consider using appropriate performance indicators such as the
monthly trends in - the number of out of calibration MMD’s; or
the number of MMD’s past their calibration due date; number of
MMD’s being used and not controlled; etc. Use these indicators
to tighten and improve the effectiveness of your MMD process.
If you have significant laboratory activities, it might be
useful to establish similar performance indicators to measure
the effectiveness of laboratory controls.
Þ
Clause 7.6 does not require a ‘documented’ procedure. However,
you must identify and document all processes addressing this
clause as part of your QMS (see clause 4.1). For these
processes, you must also identify what specific documents,
controls and resources are needed (see clause 4.2.1d. and
7.1b.). You could use a product quality plan; documented
procedure or other combination of specific practices,
procedures, documents and methods. Look at the risks related to
your product, processes and resources in determining the extent
of documented controls you need to have (also see clause 4.2.1
notes).
Þ
Performance indicators (to measure the effectiveness of
processes that control MMD’s) may include reduction - in MMD’s
found past due for calibration and being used; reduction in
damaged, uncontrolled and uncalibrated MMD’s being used;
reduction in untrained personnel found using MMD’s; reduction
in lost MMD’s; reduction in MMD’s found out of calibration; no
uncertified external laboratories being used;
etc.
|